Low temperature constant bath 200 300Version 1.1z Thank you very much for purchasing the product of Yamato ScientificCo., Ltd..z Read this operating m
8Preparation and confirmation before using1. Pour the liquid into the reservoir The amount of the overflow liquid changesdepending on the shape and t
9Method of operationBU2001. Turn the power supply switch on The power supply turns on when the powersupply switch is moved up.2. Set the setpoint (de
10BU3001. Turn the power supply switch on The power supply turns on when the powersupply switch is moved up.2. Set the setpoint (desired temperature)
11Method of maintenance CautionWhen maintaining, please pull out the power supply plug from the outlet for safety. Do notdecompose the equipment.1. M
123. Refrigerator Please detach the cover of the front of this machine regularly and remove the garbage which adheresto the condenser fin with a clea
13After-sales service and guarantee1. When you request the repair2. Guarantee book Please record the error code, turn off the powersupply switch, and
14Specification tableModelBU200 BU300Performance ↓↓↓↓Temperature Range (°C)−10 80 to − 20 80 toCooling Capacity (J/s)350(300 kcal/h)1 at 20 °C290 (
15Chart of connecting wiresBU200BU300 8 Applicable to BU200 onlySymbol NameFG Frame groundELB Eart
16Exchange part tableBU200Symbol Part name Code number Specification MakerELBEarth leakage circuit-breaker 2-06-005-0019 AC100V 15A BJS1531 Matsushita
Table of ContentsAttentionAttentionAttentionAttention_________________________________________________________________________________________________
1AttentionFor the important warnings, the following symbols are used depending on thelevel and content of danger. Warning For personal
2MeritBU200 and BU300 have following merits1. CFC-Free refrigeration system2. Featuring earth leakage circuit-breaker3. Building up optimized heati
3Please observe the following rulesInstallation Warning1. Please take the earthPlease take the earth to avoid the electric shock hazard by the leak.
4When using Warning1. Do not sprinkle water on the main body When water splashes to the electricalcomponent, the leak and the electric shock arecaus
53. About the liquid in the reservoir◎◎◎◎○○○○×××× Please use water or the ethyl alcohol solution.(1) When using water(1) When using water(1) When usi
6Name of each part and functionBU200Attached goodsauxiliary printed matters • Instruction manual • Guarantee bookDrain entrance()page11-2.)Power suppl
7BU300Attachedgoodsauxiliary printed matters • Instruction manual • Guarantee book• Bath cover complete set for low temperatureDrain entrance()page11-