Freeze Dryer
Model
DC400/800
Instruction Manual
- Second Edition -
Yamato Scientific Co. LTD.
z Thank you for purchasing "Freeze Dryer, DC
Series" of Yamato Scientific Co., Ltd.
z To use this unit properly, read this "Instruction
Manual" thoroughly before using this unit.
Keep this instruction manual around this unit for
referring at anytime.
WARNING!:
Carefully read and thoroughly understand the
important warning items described in this manual
before using this unit.
Vue de la page 0
1 2 3 4 5 6 ... 32 33

Résumé du contenu

Page 1 - DC400/800

Freeze Dryer Model DC400/800 Instruction Manual - Second Edition - Yamato Scientific Co. LTD. z Thank you for purchasing "Freeze Dryer, DC S

Page 2

6 Before Using this unit Requirements for Installation 4. Do not modify • Modification of this unit is strictly prohibited. This could cause a fa

Page 3 - Contents

7 Before Using this unit Requirements for Installation CAUTION! 6. Use specified receptacle for power source • Choose a correct power distributio

Page 4

8 Before Using this unit Installation Method 1. Unlock the stopper of the caster. Pulling up the lever of the stopper for caster releases the lock.

Page 5

9 Description and Function of Each Part Main Unit Control panelTrap bath Caster Exhaust port Power code Power switch (Earth leakage breake

Page 6 - Table of Illustrated Symbols

10 Description and Function of Each Part Control Panel TROUBLEDEFROSTONOFFONOFFPUMPON→STARTOFFREFRIGE Part Name Function ① Refrigerator S

Page 7

11 Operation Method Pre-freezing At pre-freezing process, quickly and completely freeze to the inside at sufficient lower temperature than the eute

Page 8

12 Operation Method Procedure of Operation Before operation, prepare a drying chamber, a vacuum pump, a vacuum hose, a tandem tube, etc. Vacuum Pum

Page 9

13 Operation Method Procedure of Operation 4. Connection with the vacuum pump • Connect the vacuum hose to the exhaust port on the left side. (Ex

Page 10 - Before Using this unit

14 Operation Method Procedure of Operation 8. Check the vacuum gauge • Make sure that inside of the chamber attains vacuum with the vacuum gauge

Page 11 - CAUTION!

15 Operation Method Procedure of Operation 12. After freeze-drying process finished, turn “VENT” face of the vacuum valve upward • Turn “VENT” fa

Page 13 - Main Unit

16 Operation Method Defrost Operation 1. Remove the chamber for flask installation • Remove the chamber for flask installation. If ice coating i

Page 14 - Control Panel

17 Handling Precautions WARNING! If a problem occurs If smoke or strange odor should come out of this unit for some reason, turn off the power ke

Page 15 - Operation Method

18 Handling Precautions Upper limit of inside bath temperature Do not operate continuously when the temperature of the bath is higher than -20℃.

Page 16

19 Maintenance Method Daily Inspection and Maintenance WARNING! • Disconnect the power cable from the power source when doing an inspection or ma

Page 17

20 Long storage and disposal When not using this unit for long term / When disposing CAUTION! When not using this unit for long term… • Turn off t

Page 18

21 In the Event of Failure… Trouble Shooting Condition Possible Causes The device does not start when turning on the power switch. • Power plug i

Page 19

22 After Service and Warranty In Case of Request for Repair If the failure occurs, stop the operation, turn OFF the power switch, and unplug the

Page 20

23 Specification Main Unit Model DC400 DC800 Method Storage water heater body cooling method Dehumidifying capacity Max. 0.6kg Max. 1.0kg Lowes

Page 21 - WARNING!

24 Specification Optional Accessories Name Specification Product codeMounting cock Inner dia.: 18.5mm Mounting cock pitch 96mm Number of port 12

Page 22

25 Wiring Diagram DC400 1PELBFM234AC100VCNPAMV3MV1X6X1VAC3123RSCOVRC1PCTC2RFX1-2 X1-3 X6VACsw1VACsw2TH85X1-17422a101178NONO9626X5-115X2-117X5-246X4

Page 23 - Maintenance Method

Contents  Cautions in Using with Safety...1 • Explanation...

Page 24 - Long storage and disposal

26 Wiring Diagram DC800 Timer (Refrigerator delay) Start relay Overload relay Operation condenser Start condenser Refrigerator T3 PTC OVR1, 2 C1

Page 25 - In the Event of Failure…

27 Replacement Parts Table Common Parts Symbol Part Name Specification Manufacturer Code No. MV1 Solenoid valve (Refrigerator) NEV-603DXF Sagin

Page 26 - After Service and Warranty

28 Reference List of Dangerous Substances Never use explosive substances, flammable substances and substances that include explosive or flammable

Page 27 - Specification

29 Responsibility Please follow the instructions in this document when using this unit. Yamato Scientific has no responsibility for the accidents o

Page 29 - Wiring Diagram

1 Cautions in Using with Safety Explanation MEANING OF ILLUSTRATED SYMBOLSVarious symbols are used in this safety manual in order to use the unit

Page 30

2 Cautions in Using with Safety Table of Illustrated Symbols Warning Warning, generally Warning, high voltage Warning, high temperatureWarnin

Page 31 - Replacement Parts Table

3 Cautions in Using with Safety Fundamental Matters of “WARNING!” and “CAUTION!” WARNING! Do not use this unit in an area where there is flammabl

Page 32 - Reference

4 Cautions in Using with Safety Fundamental Matters of “WARNING!” and “CAUTION!” CAUTION! During a thunder storm During a thunderstorm, turn off

Page 33 - Yamato Scientific Co., Ltd

5 Before Using this unit Requirements for Installation WARNING! 1. Always ground this unit • Connect the power plug to a receptacle with ground