ROTARY EVAPORATORRE210RE510First Edition● Thank you for your Yamato Scientific RE Series Rotary Evaporator purchase.● For best test date, we recommend
8Identification of Partswith Condenser CRE210 (type C)* Please understand that our products are subject to some specification changes without notice.*
94. Installation/AssemblyRE210 1. Set the stand at a stable place.Unpack the package and set the stand of the body on a stable place. If you do not
10 Installation/AssemblyRE210 4. Attach the motor to the motor support clamp.1. Fit the motor bearing bar to the motor support clamp, and fasten th
11Installation/Assembly RE510 6. Set the body at a stable place.Be sure to set the body at a stable place. 7. Attach the motor to the body.1. In
12Installation/Assembly RE210 and RE510 Steam Duct 8. Inserting the steam duct into the motor1. Before installing the steam duct check that the O-
13Installation/AssemblyRE210 and RE5109. Set the cooling condenser nut, coil ring and vacuum seal to the condenseror condenser connector.1.Remove the
14 Installation/AssemblyRE210 11. Connecting the cooling condenser holder shaft (glass set B and C only).Fit the cooling condenser holder shaft fir
15 Installation/AssemblyRE510 13. Connecting the cooling condenser holder shaft (glass set B & C only)Fit the cooling condenser holder shaft fi
16 Installation/AssemblyRE210 and RE510 15. Connecting the Evaporation and Receiving Flasks.●●●● Evaporation Flask1. Turn the blue flask remover
17Installation/AssemblyArm Jack for RE21017. Arm jack JK 200(sold separately and exclusively used for RE210)●●●● To set at a certain height1. Turn
Table of Contents1. Specifications...
18Installation/AssemblyLift for RE51018. Manual lift (Supplemental function of RE510)Adjust the lift by using the two knobs located on the right side
19Installation/AssemblyConnecting the Vacuum Hose and Water Supply19. Connecting the Vacuum Hose and Water Supply1. Remove plastic threaded joints.2.
20Installation/AssemblyPower Requirements 20. Wire Connection1. Connect the motor cable to the back of the main body.2. Connect the power cord of the
215. How To Operate 1. Cooling Condenser● Glass A or B Circulate the cooling water or alcohol in the cooling condenser.● Glass C Put dry ice and p
226. Troubleshooting GuideProblem Solving ChartTrouble & CountermeasureCheck the following points if trouble occurs. Contact Yamato’s Technical Se
237. Wiring DiagramRE210Symbol Name of PartsT1 TransformerSW1 Power SwitchSPC1 Speed Control PackM1 MotorG TachogeneratorC1 Motor CondenserVR1 Resiste
24Wiring DiagramRE510Symbol Name of PartsT1 TransformerSW1 Power SwitchSPC1 Speed Control PackM1 MotorG TachogeneratorC1 Motor CondenserVR1 Resister t
258. Replacement PartsName of Parts Parts No. ApplicationCooling Condenser (A) RG00A-30021 For A typeCooling Condenser (B) RG00B-30020 For B typeCooli
269. After Sale Service and WarrantyRequest for RepairWhen you request repairIf any troubles should occur, stop the operation immediately, turn the po
11. SpecificationsType RE210 RE510Rotation Speed *120-180rpmAccuracy of display rotation speed *1-±3rpm(at 20-180rpm)Drive system Worm gear systemRo
22. Safety InformationSafety SymbolsSafety InformationThis instruction manual and our products apply various indications for safety. Ignoring theseind
3 Safety InformationSafety PrecautionsIf the motor overloads - Stop operation immediately.If you continue operation under abnormal conditions, the mot
4Safety InformationSafety PrecautionsCleaning the exterior of the RE series evaporatorDo not use any volatile chemicals to clean the exterior of this
5Safety InformationHazardous MaterialDo not use the unit in flammable or explosive gas environments of substanceslisted below. Do not evaporate explos
63. Identification of Partswith Condenser ARE210 (type A)StandControl KnobVacuum PipeOpening to PourSampleCooling WaterPipeCockCooling Condenser (type
7Identification of Partswith Condenser BRE210 (type B)Cooling Condenser HolderCooling Condenser Holder ShaftCooling Condenser (type B)Condenser Conne