ROTARY EVAPORATORRE440RE540(100V)First Edition● Thank you for your Yamato Scientific RE series Rotary Evaporator purchase● For best test data, we reco
7Identification of PartsWith Condenser CRE440 (type C)RE540 (type C)* Please understand that our products are subject to some specification changes wi
84. Installation/Assembly1. Set the Body at a stable place.Be sure to set the body at a stable place.2. Fix the motor to the body.1. Insert the motor
9Installation/Assembly 3. Inserting the steam duct into the motor1.Before you insert the steam duct, make sure the O-rings inside the motor center
10Installation/Assembly4. Set the cooling condenser nut, coil ring and vacuum seal to the condenser orcondenser connector.1. Remove the cooling conden
11Installation/Assembly6. Connecting the cooling condenser holder shaft. (glass set B & C only)Fit the cooling condenser holder shaft firmly into
12Installation/Assembly8. Connecting the Evaporation and Receiving Flasks.Insert the evaporation flask into the steam duct and hold by the flask clamp
13Installation/AssemblyRemoving the Evaporation Flask and Steam Duct 10. Evaporation flaskRemover the flask clamp while holding the evaporation flask
14Installation/AssemblyLift 12. How to operate the lift.Easily move the lift up or down by using the lift switch located on the right side of the uni
15Installation/AssemblyVacuum Hose and Water Supply13. Fit the hose joints to the water supply, drain and vacuum hoses at first, andconnect to the coo
16Installation/AssemblyPower Requirements 14. Connect the power plug into an outlet.1. Connect the power cord of the body to AC100V power source.Ne
Table of Contents1. Specifications ...
175. How To Operate 1. Pour cooling water/alcohol into the cooling condenser.●In case of glass A or B Circulate the cooling water in the cooling cond
186. Troubleshooting GuideProblem Solving ChartTrouble & CountermeasureCheck the following points if there should or seem to be some machine troub
19Wiring DiagramRE440RE440Symbol Name of Parts Symbol Name of PartsP1 Power Plug CN1 Drive SocketS1 Power Socket ( for Vacuum Controller) MC Drive Cab
20Wiring DiagramRE540RE540Symbol Name of Parts Symbol Name of PartsP1 Power Plug MC Drive CableS1 Power Socket (for Vacuum Controller) F1 Fuse (2A)SW1
218. Lists of Exchange PartsName of Parts Parts No. ApplicationCooling Condenser (A) RG00A-30021 For A typeCooling Condenser (B) RG00B-30020 For B typ
229. After Sale Service and WarrantyRequest for RepairWhen you request repairIf any troubles should occur, stop the operation immediately, turn the po
11. SpecificationsType RE440 RE540Rotation Speed *120-180rpmAccuracy of display rotation speed *1-±3rpm(at 20-180rpm)Drive system Worm gear system
12. Safety InformationSafety SymbolsGraphic indicationsThis instruction manual and our products apply various indications for safety. Ignoring thesein
2Safety InformationSafety PrecautionsIf the motor overloads - stop operation immediately.If you continue operation under abnormal overload conditions,
3Safety InformationSafety PrecautionsOnly water to be used with the Water Bath.Only water in the Water Bath. Problems can occur if anything other than
4Safety InformationHazardous MaterialDo not use the Unit in flammable or explosive gas environments of substanceslisted below. Do not evaporate explos
53. Identification of PartsWith Condenser ARE440 (type A)RE540 (type A)StandVolume KnobVacuum Regulator VR100Vacuum PipeOpening to PourSampleCooling W
6Identification of PartsWith Condenser BRE440 (type B)RE540 (type B)Connecting TubeCooling Tube HolderCooling Tube Holder ShaftCooling Tube (type B)